Images are representations only.

Cerabar PMP51B - pressure transmitter

Brand: Endress+Hauser

This transmitter belongs to the new Cerabar generation. It combines robustness with numerous benefits: e.g easiest local or remote operation and condition-based maintenance which promotes better management of scheduled plant shutdowns.
The software is designed to simplify the handling. Intuitive and clear wizards guide you through the commissioning and proof testing of the device. The transmitter can be remotely controlled via app on a mobile device and a secure Bluetooth connection.

 
  • Pressure measuring range: up to 400 bar (6000 psi)

  • Process temperatures up to 400°C (752 °F) with diaphragm seal

  • Accuracy: up to +/- 0.055%

Features
  • Reduce systematic failures - error free SIL commissioning and instrument guided proof testing

  • Reduce on site accidents and protect equipment when instruments are in difficult to reach areas

  • Wireless control of the device in the process area with the SmartBlue App without process interruption

This transmitter belongs to the new Cerabar generation. It combines robustness with numerous benefits: e.g easiest local or remote operation and condition-based maintenance which promotes better management of scheduled plant shutdowns.
The software is designed to simplify the handling. Intuitive and clear wizards guide you through the commissioning and proof testing of the device. The transmitter can be remotely controlled via app on a mobile device and a secure Bluetooth connection.

 
  • Pressure measuring range: up to 400 bar (6000 psi)

  • Process temperatures up to 400°C (752 °F) with diaphragm seal

  • Accuracy: up to +/- 0.055%

Features
  • Reduce systematic failures - error free SIL commissioning and instrument guided proof testing

  • Reduce on site accidents and protect equipment when instruments are in difficult to reach areas

  • Wireless control of the device in the process area with the SmartBlue App without process interruption

Specifications

Measuring principle

  • Absolute and gauge pressure

Characteristic

  • Pressure transmitter with metal membrane for accurate measurement in liquids or gases.
  • Available with diaphragm seals.
  • Intuitive and clear wizard for commissioning.

Supply voltage

  • Ex d, Ex e, non-Ex: 10.5...35V DC Ex i: 10.5...30V DC

Reference Accuracy

  • Standard: up to 0.075 %
  • Platinum: up to 0.055 %

Long term stability

  • 0.1% of URL/year
  • 0.2% of URL/ 5 years
  • 0.25% of URL/ 10 years

Process temperature

  • Standard: -40°C…+125°C (-40°F…+257°F)
  • Diaphragm seal: -70°C...+400°C (-x°F...+752°F)

Ambient temperature

  • -40°C...+85°C (-40°F...+185°F)

Measuring cell

  • 400 mbar...400 bar (6psi...6,000psi)

Smallest calibratable span

  • 20 mbar (0.3 psi)

Vacuum resistance

  • 50 mbar (0.725 psi)

Max. Turn down

  • 100:1

Max. overpressure limit

  • 2,800 bar (40,600 psi)

Process connection

  • Thread: G1/2...G2, R1/2, MNPT3/4...MNPT2, FNPT1/2
  • Flange: DN25...DN80, NPS 1"...3"

Process connection hygienic

  • Process connection hygienic

Material process membrane

  • 316L, AlloyC, Tantal, Monel, PTFE

Material gasket

  • PTFE

Fill fluid

  • Silicone oil
  • Inert oil
  • Vegetable oil
  • High/ low temperature oil

Material housing

  • Aluminium

Communication

  • 4...20 mA, 4...20 mA HART PROFINET with Ethernet-APL Bluetooth® wireless technology and app (optional)

Certificates / Approvals

  • ATEX
  • IEC Ex
  • CSA C/US
  • NEPSI

Safety approvals

  • SIL

Design approvals

  • EN10204-3.1
  • NACE MR0175
  • NACE MR0103

Hygienic approvals

  • Hygienic approvals

Drinking water approvals

  • NSF

Specialities

  • Bluetooth® Operation and maintenance SmartBlue App
  • RFID TAG for easy identification
  • Plug and play functionalities