Proline t-mass I 300 thermal mass flowmeter

Brand: Endress+Hauser

The patented sensor design of t-mass I provides unprecedented measurement stability in thermal insertion mass flow measurement. It compensates in real time for changes of process conditions: temperature, pressure, flow direction and gas type. Its compact transmitter offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables measurement reliability and compliant verification.

 
  • Measuring principle is characterized by a high operable flow range and direct mass flow measurement

  • Measurement of utility and process gases and gas mixtures in circular piping or rectangular ducts

Device properties

  • Insertion version for DN 80 to 1500 (3 to 60")

  • Bidirectional measurement; high measuring performance

  • Patented drift-free sensor with SIL 2

  • Compact dual-compartment housing with up to 3 I/Os

  • Backlit display with touch control and WLAN access

  • Remote display available

Features
  • Flexible, convenient programming based on 21 standard gases or freely definable gas mixtures thereof

  • High level of process control – premium measurement accuracy and repeatability

  • Reliable monitoring – detection of process disturbances and reverse flow

  • Flexible installation – suitable for large dimensional range and circular pipes or rectangular ducts

  • Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses

  • Reduced complexity and variety – freely configurable I/O functionality

  • Integrated verification – Heartbeat Technology

The patented sensor design of t-mass I provides unprecedented measurement stability in thermal insertion mass flow measurement. It compensates in real time for changes of process conditions: temperature, pressure, flow direction and gas type. Its compact transmitter offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables measurement reliability and compliant verification.

 
  • Measuring principle is characterized by a high operable flow range and direct mass flow measurement

  • Measurement of utility and process gases and gas mixtures in circular piping or rectangular ducts

Device properties

  • Insertion version for DN 80 to 1500 (3 to 60")

  • Bidirectional measurement; high measuring performance

  • Patented drift-free sensor with SIL 2

  • Compact dual-compartment housing with up to 3 I/Os

  • Backlit display with touch control and WLAN access

  • Remote display available

Features
  • Flexible, convenient programming based on 21 standard gases or freely definable gas mixtures thereof

  • High level of process control – premium measurement accuracy and repeatability

  • Reliable monitoring – detection of process disturbances and reverse flow

  • Flexible installation – suitable for large dimensional range and circular pipes or rectangular ducts

  • Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses

  • Reduced complexity and variety – freely configurable I/O functionality

  • Integrated verification – Heartbeat Technology

Specifications

Sensor Features

  • High level of process control – premium measurement accuracy and repeatability
  • Reliable monitoring – detection of process disturbances and reverse flow
  • Flexible installation – suitable for large dimensional range and circular pipes or rectangular ducts
  • Insertion version for DN 80 to 1500 (3 to 60")
  • Bidirectional measurement; high measuring performance. Patented drift-free sensor with SIL 2

Transmitter features

  • Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses
  • Reduced complexity and variety – freely configurable I/O functionality
  • Integrated verification – Heartbeat Technology
  • Compact dual-compartment housing with up to 3 I/Os
  • Backlit display with touch control and WLAN access
  • Remote display available

Nominal diameter range

  • DN 80 to 1500 (3 to 60")

Wetted materials

  • Materials for insertion tube Stainless steel, 1.4404 (316/316L)
  • Process connections, process coupling
  • Stainless steel, 1.4404 (316/316L)
  • Sensing element Unidirectional
  • Stainless steel, 1.4404 (316/316L) Alloy C22, 2.4602 (UNS N06022)
  • Bidirectional Stainless steel, 1.4404 (316/316L)
  • Reverse flow detection Stainless steel, 1.4404 (316/316L)
  • Clamping rings PEEK PVDF 1.4404 (316/316L) Flat ring seal EPDM FKM

Measured variables

  • Massflow
  • temperature
  • standard volume flow
  • volume flow
  • Free air delivery
  • velocity
  • heat flow
  • energy flow
  • density

Max. measurement error

  • Gas: 1.0% o.r. (10 to 100% o.f.s.), 0.1% o.f.s. (1 to 10% o.f.s.)

Measuring range

  • 20 to 733501 kg/h (44 to 1669340 lb/h)

Max. process pressure

  • -0.5 to 20 bar_g (-7.25 to 290 psi_g)

Medium temperature range

  • -40 °C to +180°C (-40 °F to +356°F)

Ambient temperature range

  • -40 to 60°C (-40 to 140°F)
  • Optional: Transmitter: -50 to 60°C (-50 to 140°F)
  • Sensor: -60 to 60°C (-60 to 140°F)

Transmitter housing material

  • Aluminium, AlSi10Mg, coated Polycarbonate

Degree of protection

  • IP66/67, Type 4X enclosure

Display/Operation

  • 4-line backlit display with touch control (operation from outside) Configuration via local display and operating tools possible Remote display available

Outputs

  • 3 outputs: 4-20 mA HART (active/passive) 4-20 mA (active/passive) Pulse/frequency/switch output (active/passive) Relay output

Inputs

  • Status input 4-20 mA input

Digital communication

  • HART, Modbus RS485

Power supply

  • DC 24V AC 100 to 240V

Hazardous area approvals

  • ATEX, cCSAus, IECEx, NEPSI, JPN, UK Ex, EAC, KC

Product safety

  • CE, C-tick

Functional safety

  • Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

Metrological approvals and certificates

  • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
  • Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a

Pressure approvals and certificates

  • CRN

Material certificates

  • 3.1 material NACE MR0175/MR0103